EasyManua.ls Logo

VTI Instruments EX1629 - Vtex1629_Self_Cal_Store

Default Icon
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.vtiinstruments.com
EX1629 Command Set 227
vtex1629_self_cal_store
FUNCTION PROTOTYPE
ViStatus vtex1629_self_cal_store (ViSession vi);
FUNCTION PARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
DATA ITEM RESET VALUE
Not applicable to this function.
DESCRIPTION
This function takes the current self-calibration image and stores it to nonvolatile memory, enabling it to be loaded
upon instrument power cycle and reboot. Any other previously stored self-calibration data will be lost. Since the
existence of nonvolatile self-calibration data represents a permanent (although revocable) change from the full
calibration settings, its presence is able to be queried. See the vtex1629_self_cal_is_stored command.
NOTE Performing a self-calibration does not automatically store the determined calibration constants in
nonvolatile memory. However, it does turn the determined constants into the current self-
calibration data. A call to this function will store that image permanently in non-volatile storage.
EXAMPLE
ViSession instrumentHandle;
ViStatus status;
status = vtex1629_self_cal_store(instrumentHandle);