www.vtiinstruments.com
EX1629 Command Set 297
vtex1629_soft_trig
FUNCTION PROTOTYPE
ViStatus vtex1629_soft_trig (ViSession vi);
FUNCTION PARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
DATA ITEM RESET VALUE
Not applicable to this function.
DESCRIPTION
This function sends a software-generated trigger event to the EX1629.
A software trigger is only legal when the trigger source is configured as pattern (see vtex1629_set_trigger_source).
EXAMPLE
ViSession instrumentHandle;
ViStatus status;
…
status = vtex1629_soft_trig(instrumentHandle);