220 EX1629 Command Set
vtex1629_self_cal_clear_stored
FUNCTION PROTOTYPE
ViStatus vtex1629_self_cal_clear_stored (ViSession vi);
FUNCTION PARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
DATA ITEM RESET VALUE
Not applicable to this function.
DESCRIPTION
This function clears self-calibration data from nonvolatile memory. It does not, however, affect the current self-
calibration data, regardless of whether it came from a recent self-calibration or a loading from nonvolatile memory
upon power-up initialization. Thus, the effective calibration will not be changed by this call. To return to an
operating state where only the full calibration is used, the vtex1629_self_cal_clear function can be called, or the
system can be rebooted.
EXAMPLE
ViSession instrumentHandle;
ViStatus status;
status = vtex1629_self_cal_clear_stored (instrumentHandle);