www.vtiinstruments.com
EX1629 Preparation for Use 35
Wideband Output
TEDS
COMM
External customer
shunt cal
resistor terminals
Internal
Rcal
55K
-R Cal
+R Cal
Pin 5
Pin 4
Strain
Bridge
8-pin telcom
connector (RJ-45)
350
120
User
10K
10K
+ Excitation
- Excitation
(1 of 48 channels)
+Excitation
+Sense
-Sense
+Excitation Sense
-Excitation Sense
-Excitation
+Excitation Sense
-Excitation Sense
RJ-45 Pin#
12345678
Pin 2
Pin 7
Pin 6
Pin 3
Pin 8
Pin 1
Wideband Output
10K
10K
Instrumentation
Amplifier
+
-
x1
x10
x100
Hi
Lo
Shield
150
150
-Sense
-Cal
AGND
+Sense
Wagner
+Cal
AGND
430 kHz
LOW-PASS
FILTER
60.3 kHz
LOW-PASS
FILTER
ADC
TRIFILAR
FILTER
FIGURE 2-8: WIDEBAND OUTPUT DIAGRAM
DRIVER INSTALLATION
The EX1629 is shipped with a VTI Instruments Corp. Drivers and Product Manuals CD which
includes software drivers, user’s manuals, the VTI Instruments Product Catalog, as well as some
third-party software. Please refer to the ReadMe.txt file included on the Distribution CD for
installation instructions specific to revision of the CD provided.
NETWORK CONFIGURATION
With its default network configuration, the EX1629 will attempt to locate a DHCP server. If one is
found, the IP address assigned by the DHCP server will be used. Otherwise, after a timeout of 20
seconds, the unit will attempt to obtain an IP address by using AutoIP.
NOTE At any time, the EX1629 can be returned to a known, default network configuration by using the
LCI (LAN Configuration Initialize) mechanism. See Reset Button for more information.