EX1629 Introduction 19
55 kΩ (0.1%, 25 ppm/°C) per channel standard (0.05%, 5 ppm/°C optional)
Front panel connection shared among 16 channels
Connection resistance: 16 Ω typical
Software selectable: local (across completion resistor) or remote
QUARTER-BRIDGE STRAIN MEASUREMENTS
Note 1: Nominal for balanced bridge
Note 2: Conditions:
GF = 2.0, R
comp
= 350 Ω, balanced excitation
<30 days, ±5 ºC from last self-calibration
15 ºC to 35 ºC, 1 year from full calibration
Assumes the excitation voltage is measured and used in the conversion. Valid for 30 days, ±5 ºC.
Includes the stability effects of the excitation source.
60 minute warm-up
Exclusive of lead wire desensitization errors
Exclusive of gage errors
Exclusive of noise
Note 3: <30 days, ±5 ºC from unstrained voltage measurement.
Note 4: Only applies outside of self-calibration window
FULL-BRIDGE STRAIN MEASUREMENTS
Note 1: Nominal for balanced bridge
Note 2: Conditions:
GF = 2.0, balanced excitation, remote sense
<30 days, ±5 ºC from last self-calibration
15 ºC to 35 ºC, 1 year from full calibration
Assumes the excitation voltage is measured and used in the conversion. Valid for 30 days, ±5 ºC.
Includes the stability effects of the excitation source.
60 minute warm-up
Exclusive of gage errors
Exclusive of noise
Note 3: <30 days, ±5 ºC from unstrained voltage measurement.
Note 4: Only applies outside of self-calibration window