CHAPTER 13 APPENDIX
In custom applications it should be made sure that at least 50 ms
elapse between two consecutively sent letters. The communication
has lower priority than the measuring routines and at higher rates
letters may get lost. For some commands the measuring program is
transiently stopped. Hence, data transfer should not occur during
measurements.
In addition, also the WinControl software is available, which has
been optimized for the communication between PC and MINI-PAM
and features a number of most comfortable functions for data
acquisition and analysis (see separate manual).
Command Corresponding
point in
MODE-menu
Description
? 37: Date of Software version (current
EPROM)
a1/a0 12: Act. light on/off
aix 15: Act. Intensity with setting x (x = 0
... 12)
awx 14: Act. Width with setting x (x = 10 s
... 5 min)
a01/a00 10: AUTO-OFF on/off
af or afx 16: ACT.-FACTOR (x = 0.5 ... 1.5)
ay1/ay0 13: Act. Light + YIELD on/off
b BREAK, to stop all running
functions
be1/be0 Enable/Disable beep-function
bp or bpx Beep with length x (1 = 10 ms)
c1/c0 28: CLOCK on/off
ct or ctx 30: CLOCK time (interval x = 10 ...
990 s)
ci or cix 29 CLOCK item (1..6)
d or dx 48: Damping setting (x = 1 ... 3)