13.2 List of warnings and errors
Errors in MINI-PAM performance and warnings concerning sub-
optimal use of the instrument are signalled by messages in the upper
left corner of the display line. The following list briefly describes the
various error messages:
Err. OVERFLOW: >3500
Maximal signal level was exceeded. The distance to the
sample should be increased. Alternatively, the GAIN (menu
point 49) or the MEAS-INT (menu point 50) may be
decreased. In the latter cases, the zero offset must be newly
determined (2: AUTO-ZERO).