CHAPTER 7 THE MODE-MENU
7 The Mode-Menu
The MODE-menu contains 51 items corresponding to a variety
of measured values, instrumental settings or special commands. The
positions of the various menu points were arranged for optimal
practicability, with the most frequently used functions being closest
to the standard position 1.
Increasing or decreasing position numbers are selected by the -
or -arrow keys, respectively. Changes are terminated via SET or
MODE. Starting from position 1, at increasing numbers there are
mostly MODE-points involving commands (for example, 2: AUTO-
ZERO), while at decreasing numbers the MODE-points for
instrumental settings prevail (for example, 50: MEASURING
LIGHT INTENSITY). Some of the MODE-menu positions can be
directly reached via double key operations (see list in section 4.2
above).
Irrespective of the selected menu position, a YIELD-
measurement can be initiated at any time by pressing the START-key.
Normally, the system then automatically returns to the menu position
1 where the measured data set is displayed. The only exceptions are
menu-positions 11, 25-27 and 34, where the displayed values are of
primary interest.
The operations related to the various points of the MODE-menu
are either directly carried out via SET (e.g. 2: AUTO-ZERO: 50) or
initiated/terminated (e.g. 50: MEAS-INT: 8) by pressing SET.
Settings are changed by arrow key operations (, ) and become
immediately effective. The numbers following the double points
show the present settings.