CHAPTER 6 DESCRIPTION OF THE MEMORY-FUNCTION
(MODE-menu point 25) had been carried out beforehand (see
12.3.4).
Further operation of SET (2x) leads back to the original display
with time and date.
Using the arrow keys ∧ and ∨ one can move within the memory
and display any previously recorded data sets.
All data stored in MEMORY can be cleared by the CLEAR
MEMORY function (MODE-menu point 39). For safety's sake, this
command does not only require execution by SET, but in addition
confirmation by the ∧-key. The memory is organized in form of a
ring storage and its clearance normally is not required, as old data
will be automatically overwritten.
The MEMORY-front normally corresponds to the MEM-No.
under which the last set of data was stored. It can be moved to any
number between 1 and 4000 with the help of MODE-menu point 38.
After any change in instrumental settings, the complete set of
settings will be stored upon the next YIELD-measurement in the
Report-file of the WinControl program (see separate manual). This is
indicated by "Saved Settings" in the MEMORY-display.
Data stored in MEMORY can be readily transferred to a PC via
the RS 232 cable (see section 9).
13