5. Set METER SELECT to V
c
+h IN
6. Set CURRENT HOLD toggle to positive (+)
7. Adjust CURRENT HOLD control until METER reads 50 mV.
8. Set METER SELECT to V
m
.
9. Verify that METER reads 0 mV. (e.g. CURRENT HOLD command is disabled in I
o
mode.)
10. Set MODE toggle to I
c.
11. Set METER SELECT to I
m
.
12. Adjust CURRENT HOLD control until METER reads 500 pA.
13. Set METER SELECT to V
m
.
14. Verify that meter reads 5 1 mV (corresponding to the holding potential required to
pass a 500 pA current through a 10 M resister).
15. Return amplifier to initial settings.
This completes the functional checkout of the instrument. Please contact our technical support
staff if you have any questions or experience any problems.