EasyManua.ls Logo

Warner PC-505B - Whole-Cell Recording: Series R Compensation; Series R Compensation Procedure

Warner PC-505B
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PC-505B Manual, Rev. 201013
Warner Instruments
A Harvard Apparatus Company
33
Rupture is signaled by a sudden large increase in
the TEST PULSE current when in voltage clamp mode,
or as a sudden decrease in TEST PULSE voltage at the
V
m
x10 terminal when in current clamp mode. Once
the membrane is ruptured, immediately apply a
resting holding potential appropriate to the cell under
study.
An example of a ruptured membrane signal is
shown to the right.
Whole-cell recording - Series R compensation
In single channel recording, the pipette electrode
resistance (RP) is negligible compared to the patch
and seal resistances. However, in whole cell recording
this electrode resistance becomes a significant factor
and will slow the response of the amplifer.
Series R compensation is adjusted by using the SERIES R and C-SLOW controls. The SERIES
R control compensates for the resistance of the pipette opening at the cell and the C-SLOW
control compensates for the cell membrane capacitance.
A good adjustment for Series R is important
but can be a subtle operation. A basic approach is
outlined below.
Verify that the TEST PULSE is active and set to
x0.01. The trace should appear like the example
to the right.
If youve previously measured the pipette
resistance (RP), then initially set the SERIES R
dial to this value. NOTE: The SERIES R control is calibrated to 10 M/turn, so if your
RP=4.5 M, then set the dial to 0.45.
If you do not know the pipette resistance, then
set the SERIES R control to 5.0.
Next, turn on C-SLOW and adjust up until a
slight overshoot appears on the leading edge of
the trace. An example is shown to the right.

Table of Contents