EasyManua.ls Logo

Warner PC-505B - Checkout Procedures - Compensation and Current Clamp; Capacitance Compensation Checkout; Current Clamp Checkout

Warner PC-505B
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PC-505B Manual, Rev. 201013
Warner Instruments
A Harvard Apparatus Company
24
10. Set GAIN to 10 mV/pA.
11. Verify that oscilloscope reads +1 V (1 div at these settings)
12. Set VOLTAGE HOLD to negative ().
13. Verify that METER reads 10 pA.
14. Verify that oscilloscope reads 1 V (1 div at these settings)
15. Return amplifier to initial settings
Cap Comp checkout
1. Connect model cell to headstage in patch mode.
2. Set PROBE RESISTOR to high
3. Set VOLTAGE HOLD control to off.
4. Set scope to 5 V/div.
5. Set COMMAND SELECT to internal command.
6. Set COMMAND SENSITIVITY toggle to on.
7. Scope should now display large amplitude positive and negative pulses. Clipping LED
will be lit.
8. Using FAST CAP COMP controls, minimize amplitude and duration of pulses. (Leading edge
transients will remain due to characteristics of model cell.)
9. Set PROBE RESISTER to low.
10. Connect model cell to headstage in whole cell mode.
11. Turn CSLOW toggle switch on.
12. Adjust CSLOW and SERIES R controls until leading edge of square wave transitions
smoothly from one peak to the other. Full transition should take approximately 1 ms.
Remaining leading edge transients can be minimized by small adjustments to the FAST
CAP COMP controls.
13. Turn % CORRECTION toggle switch on. Verify that control decreases transition time
without increasing amplitude of spike.
14. Turn % CORRECTION off.
15. Turn LEAK SUBTRACTION on.
16. Verify that control adjusts amplitude of square wave.
17. Remove model cell from headstage.
18. Return amplifier to initial settings.
Current clamp checkout
1. Attach model cell to headstage in bath mode.
2. Set MODE toggle to I
o.
3. Set METER SELECT to V
m
.
4. Using JUNCTION ZERO control adjust METER to 0 mV.

Table of Contents