LC-201B Headstage
(50 G/500 )
single channel currents to 200 pA
whole cell currents to 20 nA.
HC-202B Headstage
(50 G/50 )
single channel currents to 200 pA
whole cell currents to 200 nA
HB-205B Bilayer Headstage
(50 G modified)
for artificial bilayer capacitances up to 250 pF
currents to 200 pA
Noise (referred to input)
Measured with an 8-pole Bessel
filter, input open, 50 G resistor:
DC to 1 k Hz 0.038 pA RMS
DC to 5 k Hz 0.170 pA RMS
± 10 V maximum, AC or DC
applied to input
± 200 mV with 10-turn control
± 100 mV with 10-turn control
1 Volt 50/60 Hz (line frequency) square wave
attenuated by COMMAND SENSITIVITY
attenuates at x0.1, x0.01, and x0.001
1.0 Volt Pulse
adjustable duration from 0.1 to 10 ms
±1000 pA maximum with Command Sensitivity @ x0.1
±100 pA maximum with Command Sensitivity @ x0.01
±10 pA maximum with Command Sensitivity @ x0.001
± 1 nA with 10-turn control
1 nA 50/60 Hz square wave
attenuates at x0.1, x0.01, and x0.001
0.33 to 8.5 s
0 to 15 pF
0 to 100 pF with 10–turn control
0 to 100 M with 10–turn control
50 G Headstage Resistor: to 50 G
500 M Headstage Resistor: to 500 M
50 M Headstage Resistor: to 50