Preparing the mass spectrometer for operation with the APGC 2-11
5. Remove the blanking plug from the corona pin mounting socket.
6. Ensure that the source isolation valve is closed.
7. Connect the APGC corona pin extension cable to the corona pin
mounting socket.
8. Pull out the locking plate on the APGC ion chamber to expose the 2
captive securing screws.
9. Locate the APGC ion chamber on the ion block and use the combined,
2.5-mm Allen wrench and cone extraction tool to tighten the 2 captive
securing screws.
10. Push the locking plate back into position.
Fitting the APGC ion chamber:
Securing screw
APGC ion chamber
APGC ion chamber cable
connector
APGC corona pin
extension cable
Corona pin mounting
socket