32 EN WM 100 TD
WM 67841c 03/2015
4 Preparation
3. Plug the free end of the power supply cable in the power
socket.
The power supply unit adjusts to the line voltage (110 V or
240 V) automatically.
The LED on the power supply unit lights up green.
Result The power supply is connected.
The therapy device is switched on and in Standby mode.
4.2.2 Connecting up the respiration hose
1. Connect the respiration hose to the device outlet.
If you want to disconnect the therapy device from the power
supply, press the clip on the connector and pull the connector out.
Do not pull on the power supply cable.
Risk of injury due to contaminated or infected patient hose
system!
A patient hose system contaminated or infected due to lack of or
incorrectly performed hygienic preparation procedures can pass
contamination or infection on to the next patient and cause
injuries.
Do not reprepare disposable hose systems.
Use a bacteria filter
Hygienically prepare reusable hose systems correctly (see "7.4
Hygienic preparation of the respiration hose", page69).