66 EN WM 100 TD
WM 67841c 03/2015
7 Hygienic preparation
7 Hygienic preparation
7.1 General information
• This product may contain disposable items. Disposable
items are intended to be used only once. So use these
items only once and do not reprocess them. Reprocessing
disposable items may impair the functionality and safety of the
product and lead to unforeseeable reactions as a result of
aging, embrittlement, wear, thermal load, the effects of
chemical processes, etc.
• Wear suitable protective equipment for disinfection work.
• Please refer to the instructions for use supplied with the
disinfectant used.
• Also observe the respective instructions for use for the therapy
device, the components, and the accessories.
• The therapy device is suitable for subsequent use on further
patients following hygienic preparation by the authorized
dealer.
7.2 Cleaning intervals
Interval Action
Weekly
Clean the therapy device (see "7.3 Hygienic preparation of the therapy device",
page 67)
Clean the respiration hose (see "7.4 Hygienic preparation of the respiration
hose", page 69)
Clean the respiratory air humidifier (see "7.5 Hygienic preparation of the
respiratory air humidifier", page 70)
In clinical areas: Disinfect the respiratory air humidifier
Monthly
Clean the air filter (see "7.3.1 Cleaning the air filter (gray filter)", page 68)
If installed: Replace the (optional) pollen filter (see "7.3.2 Replacing the optional
pollen filter (white filter)", page 69)
Every 6 months Replace the air filter