WM 67841c 03/2015
WM 100 TD EN 63
6 Settings in the menu
6.2 Setting accessories parameters
The accessories parameters are used to set the use of the
accessories.
Requirement The therapy device is in Standby mode.
1. Press the menu button .
2. Press the Accessories field.
3. Configure the required settings and confirm.
softSTART time
Intervals of 5
minutes in the
range prescribed by
the physician or
authorized dealer
(e.g., 5 mins to
max. 45 mins).
Here you can set the period of time during which the ventilation
pressure increases to the therapy pressure in the scope of the
softSTART.
If it is not possible to select this function, it must be enabled by
your physician or authorized dealer.
softPAP
Off
1
2
3
Settings 1 and 2 of the softPAP breathing relief function are
intended for patients who find exhaling against high pressure
uncomfortable. The breathing relief function reduces the
pressure early during the transition to expiration, allowing you
to breathe out more easily.
Setting 3 is suitable for patients who experience respiratory
distress with a low pressure setting. The pressure is raised
slightly during inspiration.
You can select the setting for the softPAP breathing relief here
or deactivate it if you do not wish to use the function anymore.
• Setting 1: Low breathing relief
• Setting 2: Normal breathing relief
• Setting 3: Breathing relief with inhalation assistance
This function is only available in CPAP and APAP mode. If it is
not possible to select this function in one of these modes, it
must be enabled by your physician or authorized dealer.
Parameter Possible values Description
Parameter Possible values Description
Tube type
15 mm
19-22 mm
Here you select the diameter of the hose type used. If it is not
possible to select this function, it must be enabled by your
physician or authorized dealer.