7 Hygienic preparation
3. Leave the air filter to dry.
4. Replace the air filter in the holding bracket.
Result The air filter is clean.
7.3.2 Replacing the optional pollen filter (white
filter)
1. Remove the air filter.
2. Remove and dispose of the pollen filter.
3. Insert the new pollen filter in the holding bracket.
4. Replace the air filter in the holding bracket.
Result The pollen filter has been replaced.
7.4 Hygienic preparation of the
respiration hose
1. Remove the respiration hose from the therapy device.
2. Carry out hygienic preparation of the respiration hose as
specified in the following table:
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the device.
Only use the respiration hose when it is completely dry.