EasyManuals Logo

Weinmann prisma25ST User Manual

Default Icon
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
70 EN WM 100 TD
WM 67841c 03/2015
7 Hygienic preparation
3. Rinse respiration hose off with clean water.
4. Shake respiration hose out thoroughly.
5. Hang up the respiration hose and leave it to drip dry.
6. Dry the respiration hose.
Result The respiration hose is hygienically prepared.
7.5 Hygienic preparation of the
respiratory air humidifier
Requirement The respiratory air humidifier is removed from the therapy device
(see "4.3.3 Removing the respiratory air humidifier after use",
page 38).
Part Cleaning Disinfection Sterilization
Respiration
hose
With warm water
and washing-up
liquid
Immersion
disinfection
(Recommendation:
gigasept FF
®
)
Not permitted
If you use a heatable respiration hose, please observe the
instructions for use for the respiration hose.
Risk of injury from hot element!
During and shortly after operation, the element of the respiratory
air humidifier is hot and touching it can cause burns.
Allow the element to cool down completely.
Avoid touching the element.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann prisma25ST and is the answer not in the manual?

Weinmann prisma25ST Specifications

General IconGeneral
BrandWeinmann
Modelprisma25ST
CategoryHealth Care Products
LanguageEnglish

Related product manuals