EasyManua.ls Logo

Weinmann prisma25ST - Pneumatic System Diagram; Pressure Volume Curve

Default Icon
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WM 67841c 03/2015
WM 100 TD EN 93
12 Appendix
12.2 Pressure volume curve
pV curve at AV=0.5 l and f=20/min
12.3 Pneumatic system diagram
0
0.1
0.2
0.3
0.4
0.5
0.00 5.00 10.00 15.00 20.00
Volume
[l]
Pressure [hPa]
Air filter
Inlet for
ambient air
Fan
Optional
humidifier
Pressure sensor
for patient
pressure
Optional O
2
safety valve
Optional
bacteria filter
Respiration
hose,
1.8m
Exhalation
system
Patient mask
Nasal or
full face mask
O
2
pressure
source
Flow regulator