WM 67841c 03/2015
4 Preparation
4.3 Connecting optional accessories
4.3.1 Connecting the oxygen safety valve
1. Insert the oxygen safety valve Respironics No. 302418 between
the respiration hose and the device outlet.
Result The oxygen safety valve is connected.
4.3.2 Connecting the respiratory air humidifier
Filling the respiratory air humidifier
Requirement The respiratory air humidifier is removed from the therapy device
(see "4.3.3 Removing the respiratory air humidifier after use",
page38).
Risk of injury from burning oxygen!
Supplying oxygen without special safety equipment can cause
fires and injure people.
Always use an oxygen safety valve.
Observe the information on the safe handling of oxygen.
Observe the instructions for use for the oxygen safety valve and
the oxygen supply unit.
Material damage due to overfilling!
Any escaping water can enter the device and damage it.
Remove the respiratory air humidifier from the device before
filling.
Only fill the respiratory air humidifier up to the max mark.

Questions and Answers: