46 EN WM 100 TD
WM 67841c 03/2015
5 Operation
5.3 Switching off the therapy device
1. To save energy during the day, keep the on/off button
depressed for 3 seconds.
or
If the automatic energy saving function is activated: the
therapy device switches to the Energy saving mode
automatically 15 minutes after the user has performed the last
action.
Result The therapy device is in the Energy saving mode.
5.4 Starting the therapy
1. Connect the components (see "6.1 Setting comfort
parameters", page62).
2. Connect the power supply (see "4.2.1 Connecting up the
power supply", page30).
3. If the display remains off, the therapy device is in Energy
saving mode: Press the on/off button .
The therapy device switches to the Standby mode.
4. Press the on/off button .
or
If the autoSTART-STOP function is activated: breathe into the
mask.
The automatic energy saving function can be activated in the
menu Main menu | Device | Energy saving (see "6.4 Setting device
parameters", page65).
The therapy device does not switch to the Energy saving mode
automatically if:
there is a therapy in progress;
there is a USB cable inserted;
data are being exported;
a message appears on the display.
You can activate the autoSTART-STOP function in the menu Main
menu | Comfort | autoSTART-STOP (see "6.1 Setting comfort
parameters", page62).