68 EN WM 100 TD
WM 67841c 03/2015
7 Hygienic preparation
5. If present: Connect the respiratory air humidifier up to the
therapy device (see "4.3.2 Connecting the respiratory air
humidifier", page 34).
6. Reconnect to power supply.
7. Perform a function check (see "8 Function check", page 76).
Result The therapy device and the components are hygienically prepared.
7.3.1 Cleaning the air filter (gray filter)
1. Remove the air filter.
2. Clean the air filter under running water.
Part Cleaning Disinfection Sterilization
Housing
Wipe with a damp
cloth: use water or
mild soap
Wipe disinfection
(Recommendation:
terralin
®
protect or
perform advanced
Alcohol EP)
Not permitted
High-gloss
surfaces
on the housing
Wipe with a damp
cloth: use water or
mild soap; do not use
microfiber cloths
Power supply
unit
Wipe with a damp
cloth: use water or
mild soap
Power supply
cable
Wipe with a damp
cloth: use water or
mild soap