42 EN WM 100 TD
WM 67841c 03/2015
4 Preparation
5. Fill fresh, cold water up to the marking on the underside (max.
400 ml) through the inlet.
6. Set the respiratory air humidifier upright again carefully.
7. Install the respiratory air humidifier on the therapy device
(see " Installing the respiratory air humidifier", page37).
8. Start the therapy again (see "5.4 Starting the therapy",
page46).
Result The respiratory air humidifier is filled.
Material damage due to hot water and aromatic additives!
Hot water or aromatic additives (e.g., eucalyptus oil) can damage
the housing of the respiratory air humidifier and the element.
Do not fill with hot water.
Do not use any aromatic additives.
Sterile or boiled water is only required in exceptional medical cases
when using this device at home. Do not use distilled water for
technical purposes, as it could contain microbiological pollution.