WM 67841c 03/2015
WM 100 TD EN 53
5 Operation
5.8.3 Setting the humidifier level
Requirement • The therapy device is in the Standby or Therapy mode.
• The respiratory air humidifier is filled with water (see " Filling
the respiratory air humidifier", page34).
• The respiratory air humidifier is connected to the therapy
device (see " Installing the respiratory air humidifier",
page37).
• The respiratory air humidifier is switched on (see "5.8.1
Switching on the respiratory air humidifier", page51).
The humidifier button is green and the humidifier level is
shown .
1. The and buttons can be used to increase or decrease
the humidifier level.
Result The humidifier level is set.
There are seven humidifier levels available (1-7). The level which is
suitable for you depends on the room temperature and the
humidity. The standard setting is level 4. If you wake up with dry
airways, the heating is set too low. If there is condensation in the
respiration hose in the morning, the heating is too high.
To reduce condensation in the respiration hose, we recommend
using a hose heating system.