82 EN WM 100 TD
WM 67841c 03/2015
9 Alarms and error messages
(206)
Error in the
prismaCONNECT module.
Remove and reconnect the prismaCONNECT module. If the
fault persists:
Contact the authorized dealer and have the
prismaCONNECT module replaced.
(702)
Device output is blocked. /
Water in therapy device.
Ensure that the respiration hose and device output are not
blocked.
If the fault persists:
• Check whether there is water in the device. To do so,
remove the respiratory air humidifier and side part and
tilt the device with the open side facing downward.
• If water comes out: Wait until all the water has escaped.
• Allow the device to dry until the message is no longer
displayed. In future, do not transport the device with
water in the respiratory air humidifier.
• If water collects in the respiration hose: Reduce the
humidifier level to avoid condensation.
All other error
codes
Problems with the
electronics
Disconnect the therapy device from the power supply and
reconnect it (see 4.2.1, p. 30).
If the message is still shown, contact the authorized dealer
and have the device and the respiratory air humidifier
checked.
Error code Cause Remedy