4. Write the Welch Allyn RMA number with the Welch Allyn address on the outside of the shipping
carton.
Recommended service intervals
To confirm that the device is functioning within the design specifications, perform periodic service as
indicated below. Customers can perform the basic functional verification and calibration procedures
referenced below by following the instructions in this manual.
Perform the following preventive maintenance quarterly to keep your scale in working order, or as
required based on usage or per your facility maintenance schedule, whichever comes first.
1. Have your service department check the calibration annually or as required. Calibration
instructions are provided in the scale service manual.
2. Inspect the cradle for cracks or loose mounting hardware. Replace or repair as necessary.
3. Inspect the scale’s enclosure for damage or loose or missing hardware. Replace or repair as
necessary.
4. Inspect the power adapter cable for abrasions or other sign of wear.
5. Do not expose the scale to excessive water or moisture.
6. Do not store the scale where heavy objects can be placed on it.
7. Replace the batteries annually or as required.
8. When storing the scale, remove the batteries from the battery area. Batteries can corrode over a
period of time. Make sure to check the batteries before putting the scale back into use.
9. Do not service or perform any maintenance while the scale is in use with a patient.