55
NOTE: Ohmmeters will not indicate a change in resistance of the load cell transducer when weights are applied to scale. This
is due to the extremely small change in resistance of the strain gauges employed (<1 ohm) and the fact that the bridge
configuration presents a constant value of resistance when measured from its terminals.
Symptoms and Solutions
Try operating the scale in both power modes; AC &
Battery. If it functions normally when using AC power
replace all 6 D cell Batteries.
Use a voltmeter or a multimeter to check continuity of
the adapter. Replace if needed.
Replace the Battery Holder Assembly.
(OHM out Load Cells) See Calibration procedure this
Service Manual.
See calibration instruction.
Side to side platform test using 10KG certified weight
Apply weight evenly on the center of the scale. Press
“REWEIGH” two times and record your weight
reading each time.
Remove weight from center and apply weight on
each side one at a time and Press “REWEIGH” two
times and record weight reading each time.
Examine your readings
Plot readings and they should be within ±0.4 lb./±0.2
kg from center reading.
The side that reads differently from all others is the
problem area. Repeat corner test if need be. Look for
obstructions and/or replace the Load Beam if necessary.
Scale not weighing correctly,
scale broken, drifting
Check Continuity of the Cable and if bad replace.
Cracked broken Cradle Assembly