Message Text Meaning
existing screening results. You cannot perform new screenings until
you restart the device.
Problem encountered. The
device needs to reboot.
The system was not able to start and has attempted to correct the
issue. Restart the device.
Un-calibrated device The system cannot read the device configuration file. Restart the
device. If the issue is not resolved, contact Welch Allyn Technical
Support.
For more information on troubleshooting or error messages, contact Welch Allyn
Technical Support: www.welchallyn.com.
Screening tips
If the pupils are too small, the vision screener displays a notification to adjust the lighting
in the room. Follow these steps to encourage the subject's pupils to dilate.
1. Dim the lights, if possible.
2. Close blinds or curtains, if possible.
3. The subject's back should be positioned to the interfering light source.
4. Avoid direct light into the subject’s eyes from:
ambient light sources
exterior light
incandescent and halogen lights (florescent light is the best testing environment,
if options are limited)
Screening Issues
Screening system messages
Issue
Recommendation For more information,
see
Message: The
screening
system is not
initialized.
The exam system did not initialize or is unresponsive. Restart
the device.
System restart
Message: Unable to
save PDF report
for current
subject
The file system may not have enough space. Export data and
clear the device history to free up additional storage space.
To delete records
Directions For Use Troubleshooting 41