Commissioning | 5 | English
Operating instructions UC-Serie | 89014383; Version 2 | 33
5.5 Configuring the transfer strip
A service technician configures the transfer strip in the Dosing menu, specifying the type of dosing device.
Detergent dosing device
32
1
Item Name Note
1 Reset to default setting
2 For dosing devices not from Winterhalter:
While machine fills
Parallel to the solenoid valve
Peristaltic pump With the “peristaltic pump” setting, set the output in ml/min.
For Winterhalter dosing device:
Fluidos
MPI/PPX
Set the dosing amount in the dosing menu.
3 Save settings
Rinse aid dosing device
32
1
Item Name Note
1 Reset to default setting
2 For dosing devices not from Winterhalter:
Parallel to the rinse pump
Parallel to the solenoid valve
Peristaltic pump With the “peristaltic pump” setting, set the output in ml/min.
For Winterhalter dosing device:
Fluidos
MPI/PPX
Set the dosing amount in the dosing menu.
3 Save settings