EasyManua.ls Logo

Wissner-Bosserhoff Latera - Removing Bed from Service

Default Icon
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product: Latera
Version: 01
20/57
Removing the Bed from Service
Remove the bed from service as follows:
Disconnect the bed from the mains.
Disconnect the ground wire.
Deactivate the battery.
Remove accessories.
To prevent damage during storage:
Pack or cover the bed and accessories.
Ensure that storage conditions are the same as the operating conditions.
8.3.1 Deactivating the Battery
To avoid damaging the bed and the environment during storage:
Deactivate the battery on the supervisor panel.
To deactivate the battery on the supervisor:
Disconnect the bed from the mains.
Disconnect the ground wire.
Activate the keypad by pressing the GO button on the supervisor.
Press the Thigh Rest Up + Thigh Rest Down + Trendelenburg Position buttons at the same time and hold
them for three seconds.
The battery is deactivated.

Related product manuals