Maintenance
Warning
Risk of injury when working on the bed!
Ensure that the bed is disconnected from the mains connection prior to assembly,
disassembly and maintenance.
Ensure that the castors are locked prior to assembly, disassembly and mainte-
nance.
A defective bed can cause injuries!
Have a defective bed repaired immediately.
If the defect cannot be repaired, do not use the bed.
Warning
Caution
Incorrect maintenance can damage the bed!
Ensure that maintenance is performed by customer service or trained hospital
technicians only.
Do not use the bed if any malfunction or defect occurs. In this case contact manu-
facturer or service organisation immediately.
NOTE wissner-bosserhoff
®
recommends attaching the maintenance plaque to the bed.
Ensure that the following maintenance work is performed every 12 months by the manufacturer or by a
qualified service organisation trained and certified by the manufacturer
Monthly maintenance
Check all movable parts for wear.
Maintenance every 3 months
Check function of brake lever.
Clean the piston shafts and lubricate with silicone oil.
Clean the bolts and hinges of the mattress platform (backrest, thigh rest, calf rest) and lubricate with
silicone oil.
Check mechanism of mattress platform extension.
Check bolts on castors and tighten them if necessary.
Maintenance every 12 months
12.3.1 Spare Parts
The product label is located on the inside of the longitudinal rail of the mattress platform frame. The product label
contains information for claims and ordering replacement parts.
Information about spare parts is available from:
Customer service
Sales
Our technical support department