Product: Latera
Version: 01
47/57
Safety Checks
Warning
Incorrect safety checks can cause injuries!
Ensure that safety checks are performed by customer service or authorised per-
sonnel (certified by the manufacturer) only.
Ensure that the safety checks are recorded in the service and maintenance log.
A defective bed can cause injuries!
Have a defective bed repaired immediately.
If the defect cannot be repaired, do not continue to use the bed.
Warning
In accordance with §6 of the Medical Devices Operator Ordinance, the operator is required to perform a technical
safety check on the hospital bed every 12 months.
The procedure for performing the safety check is stipulated in VDE 0751 or IEC 601.1.
The manufacturer will give a certificate to service organisations in which the manufacturer declares that the ser-
vice organisation is qualified to perform maintenance on wissner-bosserhoff
®
products.
wissner-bosserhoff
®
makes every reasonable effort within research, design and manufacturing of beds to ensure
that their products are fit for purpose and produced to the highest level of quality. However, wissner-bosserhoff
®
can take no responsibi- lity for any damage caused to the product or harm to patients, staff or other individuals as
a result of:
Not following the instructions for use, including warning and caution statements, provided in their user
manuals
Use of the product for other than its intended purpose as stated in any documentation provided by wissner-
bosserhoff
®
NOTE On request, the manufacturer will provide circuit diagrams, component part lists, descriptions, calibra-
tion instructions etc. for service personnel for the repair of ME equipment designated by the manufacturer as
repairable by service personnel.