EVOM3
World Precision Instruments 11
Setup Page
Press
Setup
on the main screen to open the
Setup
page.
Fig. 7—Setup page
Select Plate – Select the number of wells and the sampling pattern.
Mode Units – Choose the Units for sampling. (Options include Millivolts, Volts, Ohms
and K Ohms.)
Blank Handling – Lets you subtract the control blank from the current resistance
measurement.
Short Electrodes – Shorts V
1
-V
2.
and is used to nullify the galvanic charge in the
electrodes when the electrodes are placed in a saline solution for 4–8 hours to
remove the charge. See also "Probe Null" on page 14.
Date/Time  elet the ate and ime mat   data file
Calibrate  aliate the eletde (hi e a K Ω  eit) ee ntment
Diagnostics and Calibration" on page 21.
Probe Null  etalie the et hage n V
1
-V
2..
Store Screen te data file, nfige h the file ae named and i the ae
automatically numbered.
Plate Selection Page and Plate Preview
1. Press
Select Plate
on the
Setup
page to access the
Plate Selection
screen.
Fig. 8—Choose the number of wells and the sampling pattern.
2. Select the radio button for the number of wells in your plate.
3. Select the sampling pattern from the options in the bottom half of the window.