EVOM3
World Precision Instruments 13
2. Choose your units. Options include Millivolts, Volts, Ohms and K Ohms. If you
choose ohms, then you must also select an Ohms Range.
Fixed Current Variable Current
10 K 10 µA Auto 2 µA to 10 µA
50 K 4 µA
100 K 2 µA
3. Press
Return
.
Blank Handling Page
1. Press
Blank Handling
on the
Setup
page to subtract the control well blank from the
current resistance measurement.
Fig. 15—(Left) Blank Handling window.
Fig. 16—(Right) The main screen shows the oset underneath the reading.
 net the e int an emt ell filled ith media e
Reading
to make a
measurement.
3. Select the
Blank Enable
adi ttn t enale the et
4. Press
Return
 he eitane et h nde the eading n the main een
 the lan (et) i geate than the eitane eading, the eading ill dila
negatie nme in ed tet he aed data file eet the tatin
NOTE hi inlde id and eletde eitane
Setting the Date and Time
1. Press
Time/Date
on the
Setup
page to set the date and time format.
Fig. 17—The Set Date/Time screen lets you set the date and time and choose the display format.
2. Press
Set Date
. Set the appropriate date and press
Return
.