EVOM3
World Precision Instruments 25
Electrode Equilibration – Zero the electrodes' galvanic charge in saline.
Press the
Probe Null
button. Push the
NULL
button. Wait for the second
message stating ”Solution value 0.0” before returning. This action removes
an ltage et n the eletde hen atiate the
Enable/Disable
radio
ttn t et the eletde tential
Fig. 47—(Left) Null probe process.
Fig. 48—(Right) The potential dierence oset shows on the main screen under the reading.
Press
Return
to navigate back to the main menu.
4. Insert a USB drive into the USB port on the side of the EVOM3 and begin
measuring by pressing the foot switch or touching the
Store
button on the screen.
Tap the
Next Well
area on the right side of the screen to advance to the next well.
Repeat the sampling process.
 ne the lat ell ha een eded, the file ting eletin ntifiatin aea
Fig. 49—Warning that the last cell measurement was taken.
6. Press
Store
. A system message will report if the USB thumb drive is usable,
or it will display an error. If the USB thumb drive is not plugged in or it is not
recognized, a warning message will appear. Press
Cancel
and try another drive.
The
Store
function cannot be used on an unrecognized drive. EVOM3 should
end ith File aed ithin  end  aing file i till een ate 
seconds, press the
Return
e and e anthe  die ne a file ha een
aed, en the file n a mte t ei the ntent
NOTE he file aing tine an tae  t  end t mlete ee the
USB cautions on page 5.
 Reme the  thm die t tane the data file t  mte a a V file
(Excel).