RapidMatch
™
XI Spectrophotometer
16
error or results
appear
inaccurate.
measured is
damaged (e.g.
scratched)
procedure in the Calibration
section, or replace if damaged.
software not
communicating
(USB
connection).
Connect the interface cable
between the computer and the
instrument.
Close and restart the software
application. If this does not work,
reboot the computer.
Reset the instrument from the
Settings>Systems &
Diagnostics>Systems
Information screen.
the display do not
change color.
Make sure the light seal is lying
flat and not interfering with the
pressure sensors on the bottom
of the instrument. Refer to the
Light Seal Check and
Replacement section for
information on properly installing
a light seal.
working correctly. Put the
instrument into targeting mode
and place the instrument on a flat
surface and then lift. If one or
more of the sensor indicators in
the display do not change color
when lifted, there may be a
problem with the pressure
sensor. Contact technical
support.
installed properly or
damaged.
Refer to the Light Seal Check
and Replacement section for
information on properly installing
a light seal.