Remote Operation
SCPI Commands for Digital Interfaces
134 Operating Manual for XDC Series Power Supply
Clear
Protection
Event
To clear a protection mechanism that has tripped:
OUTP:PROT:CLE
This will clear all protection mechanisms and re-enable the output. If the condition
that caused the alarm still exists, the protection will be allowed to trip again.
View Power
Supply
Output
The following 3 commands query the voltage, current, or power being supplied at
the output terminals.
To measure the voltage output:
MEAS:VOLT?
To measure the current output:
MEAS:CURR?
To measure the power output:
MEAS:POW?
Configure
Auxiliary
Status Lines
To set up Auxiliary line A:
OUTP:AUXA:SOUR <aux-line-mnemonic>
The choices for <aux-line-mnemonic> include:
• NONE
• ON: Output On
• OFF: Output Off
• OVOL: Over-Voltage Condition
• UVOL: Under-Voltage Condition
• OCUR: Over-Current Condition
• UCUR: Under-Current Condition
• OPOW: Over-Power Condition
• UPOW: Under-Power Condition
• ACOF: AC Power had been turned off or failed
• OTEM: Over-Temperature Condition
• HTEM: High-Temperature Condition
• SPR: Sense Protection Tripped
• FOLD: Fold Protection Tripped
• CC: Constant Current Mode
• CV: Constant Voltage Mode
• CP: Constant Power Mode
• UNR: Output Unregulated
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xantrex XDC 60-100 and is the answer not in the manual?

Xantrex XDC 60-100 Specifications

General IconGeneral
BrandXantrex
ModelXDC 60-100
CategoryPower Supply
LanguageEnglish

Related product manuals