18
XCell
TM
ATF System with C410:V3 Controller
facilitate positioning and handling of the boxes within the facility. One can envision the
P-Box in proximity to the Filtration Assembly, while the E-Box positioned distant to the
Filtration Assembly, possibly mounted on a wall or a skid.
5.1.1 Filtration Assembly
This includes the Diaphragm Pump, Filter Housing, connection to the bioreactor, harvest line,
pump air inlet assembly, stand, plus all the housing accessories as specified in the part list.
P-Box:
a. System STOP Switch
b. Interconnect Cable Plug
c. Plugs for sensor inputs (4x4-20mA)
d. Connection for Diaphragm Pump
e. Connection for compressed air
f. Connection for vacuum supply
g. Pressure regulator (0 to 60psi, 0 to 4.1 bar)
h. Vacuum gauge (0 to -14psi, o to -0.95 bar)
i. Pressure gauge (0 to 60psi, 0 to 4.1 bar
(Not pictured in Figure 5) Pressure relief assembly set to 40psi, 2.8 bar
Figure 5. Pneumatic Box (P-Box) Connections
E-Box
The E-Box contains the HMI and PLC components, including:
the Siemens S7-1200 PLC. It is programmed using Siemens Step 7 Basic v13. The Operator
Interface Terminal (OIT or HMI) is a Siemens SIMATIC TP 1200 Comfort, programmed using
Siemens WinCC Advanced v13.
The E-box includes the following connections:
j. 120/220vac power cable with plug

Questions and Answers: