ATF System with C410:V3 Controller
The Alarm screen will display parameters in the following tables:
Field
Description
Time
Indicates time of Alarm
Date
Indicates Date of Alarm
Text
Describes Alarm
Button
Description
Horn Acknowledge
Press to turn off Horn
Select Highlight Warning/Alarm message and press to remove
All Primary
All Primary Screen Buttons are displayed to navigate to those
screens
Two (2) Pump Controller Status Relays are provided which energize per the following states:
XCell
TM
ATF System
States
State #
RELAY #1
RELAY #2
POWER OFF / ALARM
A
OFF
OFF
POWER ON /
STANDBY
B
OFF
ON
POWER ON /
RUNNING
C
ON
ON
POWER ON /
WARNING
D
ON
OFF
Two (2) Pump Interlock Relays are provided which energize per the following states:
HARVEST PUMP STATE
RELAY #3
RELAY #4
ACTIVE
ON
ON
NOT ACTIVE
OFF
OFF
An isolated Form C contact for each relay is provided for the end user to connect to any
remote monitoring system. The Harvest pump is activated to Run mode, only and only if
Relay #1 is ON, i.e. in Power On/ Running state #C or Power On / Warning state #D.
The following is a list of C410:V3 controller Warnings that can occur during normal operation:
• Flow Set Point cannot be reached. P-Flow Regulator (PV1) above maximum operating
setting.
• Flow Set Point cannot be reached. P-Flow Regulator (PV1) below minimum operating
setting.
• Exhaust Set Point cannot be reached. E-Flow Regulator (PV2) above maximum operating
setting.