P1.2
SELECT RESOLUTION
0.1 0.01
0.01
P1.3
STABILITY FILTER
LOW MEDIUM - HIGH
MED
P1.6
CALIBRATION DATA
-
-
P1.7
SET DUE CAL
NO HOURS - DAYS
NO
P1.8
RESET SETTINGS
YES NO
NO
P1.9
TEMPERATURE CAL
YES NO
-
P1.1 Selection of pH buffers
Access this setup to select the buffer family for performing the electrode calibration.
Calibration lines from 1 to 3 points can be made.
During calibration, press to exit and save the points calibrated up to that moment (see calibration
paragraph).
The instrument automatically recognizes 2 families of buffers (USA and NIST); in addition, the user has the
option of performing a manual calibration up to 2 points with customizable values.
USA Buffers: 1,68 - 4,01 - 7,00** - 10,01 - 12,45 (Factory setting)
NIST Buffers: 1,68 - 4,00 - 6,86** - 9,18 - 12,46
**Neutral point always requested as first
In measure mode at the bottom left of the display, a series of beakers
indicates the buffers with which the last automatic and manual
calibration was carried out.
P1.2 Resolution
Access this menu to choose the resolution needed when reading the
pH parameter:
0.1
0.01 -default-
P1.3 Stability criteria in pH measurement
To consider the reading of a value truthful, we recommend waiting for the measurement stability, indicated
by the icon . When the measurement is not stable, four red flashing bands appear on the display.
Access this menu to change the measurement stability criterion:
“LOW: choose this option to bring up the stability icon even in conditions of poor stability.
Readings included within 1.2 mV.
“MEDIUM” (default value): readings included within 0.6 mV.
“HIGH”: choose this option to display the stability icon only in conditions of high measurement stability,
readings within 0.3 mV.
P1.6 pH calibration data
Access this menu to view (select VIEW”) or print (select PRINT”) information on the last calibration
performed:
VIEW: The following screens will scroll automatically on the display:
First screen: CALIBRATION DATE and TIME and beakers indicating the buffers used.
Beacher
pH value of the buffer
Acid
< 6.5
Neutral
6.5 ~ 7.5
Basic
> 6.5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the XS Instruments 60 VioLab Series and is the answer not in the manual?

XS Instruments 60 VioLab Series Specifications

General IconGeneral
BrandXS Instruments
Model60 VioLab Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals