Second screen: OFFSET value of the electrode expressed in mV.
Third and possibly fourth screen: Slope% in the measuring range (one Slope% only if two calibration
points are performed, two Slope% if three points are performed).
Note: The instrument accepts calibrations with pH electrodes with Slope% between 80 - 120% only.
Outside this range of acceptability, the instrument does not allow to end the calibration and displays the
error message SLOPE OUT OF RANGE
PRINT: The following information are printed:
Model and Serial Number of the instrument.
Calibration Date and Time
OFFSET value expressed in mV
pH Range with relative Slope%
IMPORTANT: Make sure that the printer is connected and turned on and that the paper roll and cartridge are
correctly positioned. For more information, refer to the printer user manual.
NOTE: Use the original printer indicated by the manufacturer only. For information, contact your local distributor.
P1.7 pH calibration deadline
Access this menu to set a calibration deadline; this option is very important in GLP protocols.
By default, no calibration deadline is set. Use the directional keys to select HOURS or DAYS and access
with button . Use the directional keys to change the number that appears in the center of the
display, entering the hours or days that must elapse between two settings, and confirm with .
If a calibration deadline is set, the icon is present on the display in measure mode.
When the calibration deadline is activated, the instrument prevents further measurements.
The error icon and the icon representing the calibration deadline flash on the display.
The string “MAKE A NEW CAL invites the user to perform a new calibration of the pH sensor to be
able to work again.
Press the key to start the calibration.
P1.8 Reset of the pH parameter
If the instrument does not work perfectly or incorrect calibrations have been carried out, confirm YES with
the button to take all the parameters of the pH menu back to the default settings.
IMPORTANT: The factory reset of the parameters DOES NOT erase the stored data.
P1.9 Temperature calibration
All the instruments of this series are pre-calibrated for a correct temperature reading. However, if there is a
difference between the measured and the real temperature (usually due to a probe malfunction), it is

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the XS Instruments 60 VioLab Series and is the answer not in the manual?

XS Instruments 60 VioLab Series Specifications

General IconGeneral
BrandXS Instruments
Model60 VioLab Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals