HandyLab 100 Operation
35
ba77062d02 12/2017
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.
1 In the measured value display, select the measured parameter
pH or mV with <M>.
2 Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage dis-
play).
3 Thoroughly rinse the electrode with deionized water.
4 Immerse the electrode in buffer solution 1.
5 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <><>.
6 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
7 Wait for the end of the measurement with stability control or
accept the calibration value with <ENTER>.
pH 1
-180.0
24.8 °C
AutoCal TEC
15.03.2014
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
15.03.2014
08:00
AR
Buffer
mV