HandyLab 100 Operation
35
ba77062d02 12/2017
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.
1 In the measured value display, select the measured parameter
pH or mV with <M>.
2 Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage dis-
play).
3 Thoroughly rinse the electrode with deionized water.
4 Immerse the electrode in buffer solution 1.
5 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <><>.
6 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
7 Wait for the end of the measurement with stability control or
accept the calibration value with <ENTER>.
pH 1
-180.0
24.8 °C
AutoCal TEC
15.03.2014
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
15.03.2014
08:00
AR
Buffer
mV

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem SI Analytics HandyLab 100 and is the answer not in the manual?

Xylem SI Analytics HandyLab 100 Specifications

General IconGeneral
BrandXylem
ModelSI Analytics HandyLab 100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals