Operation HandyLab 100
36
ba77062d02 12/2017
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Continuing with two-
point calibration
Continuing with three-
to five-point calibration
8 If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
9 Thoroughly rinse the electrode with deionized water.
10 Immerse the electrode in buffer solution 2.
11 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <><>.
12 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
13 Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
14 If necessary, finish the calibration procedure as a two-point cal-
ibration with <M>.
The calibration record is displayed.
pH
0.0
24.8 °C
AutoCal TEC
7.000
15.03.2014
08:00
AR
Buffer
mV
15 Thoroughly rinse the electrode with deionized water.
16 Immerse the electrode in the next buffer solution.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem SI Analytics HandyLab 100 and is the answer not in the manual?

Xylem SI Analytics HandyLab 100 Specifications

General IconGeneral
BrandXylem
ModelSI Analytics HandyLab 100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals