Operation HandyLab 100
38
ba77062d02 12/2017
1 In the measured value display, select the measured parameter
pH or mV with <M>.
2 Start the calibration with <CAL>.
The calibration display appears.
3 Thoroughly rinse the electrode with deionized water.
4 Immerse the electrode in buffer solution 1.
5 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <><>.
6 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
7 Wait for the end of the measurement with stability control or
accept the calibration value with <ENTER>.
The calibration display for the next buffer appears.