HandyLab 100 Operation
39
ba77062d02 12/2017
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Continuing with two-
point calibration
8 Set the nominal buffer value for the measured temperature with
<><>.
9 Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
10 If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
pH
7.000
24.8 °C
VariCal
1
15.03.2014
08:00
Buffer
11 Thoroughly rinse the electrode with deionized water.
12 Immerse the electrode in buffer solution 2.
13 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <><>.
14 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
15 Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears.