EasyManua.ls Logo

Xylem SI Analytics HandyLab 100 - Page 39

Xylem SI Analytics HandyLab 100
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HandyLab 100 Operation
39
ba77062d02 12/2017
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Continuing with two-
point calibration
8 Set the nominal buffer value for the measured temperature with
<><>.
9 Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
10 If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
pH
7.000
24.8 °C
VariCal
1
15.03.2014
08:00
Buffer
11 Thoroughly rinse the electrode with deionized water.
12 Immerse the electrode in buffer solution 2.
13 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <><>.
14 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
15 Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears.

Related product manuals