Operation with OxiTop
®
-IDS measuring heads Multi 3630/3620 IDS
22 ba77045d02 03/2020
6
Check and, if necessary, change the settings in the Set menu.
Measuring mode
- BOD measurement (BOD standard / BOD special) or
- pressure measurement (Pressure p)
Depending on the selected measuring mode, further special set-
tings for the sample are possible.
Set name
The current date with an index (consecutive number) is automati-
cally preselected as name for the set. The name can be changed
freely (max. 8 characters [numerals / letters])
Duration
Make sure that the remaining lifetime of the battery of the measuring
head is sufficient for the specified duration. You can insert a new
battery or make use of the energy saving settings, e.g. switch off the
LED or switch on the Eco mode (radio).
Settings
7
Open the Settings / menu.
8
Check and, if necessary, change the settings for the set in the Settings
menu (see section 4.3 S
ETTINGS).
Measuring mode BOD special
Set name 190228-1
Duration 17 d
Settings
i
BOD5 End date 23.02.2019
AutoTemp, GLP
V_P 432 ml
79 Free meas. heads
Set
15.03.2020 08:00
Sample
GLP on
AutoTemp on
Alarm diff. pressure 500 hPa
Man. temperature 22°C
LED on
Number of measured values Standard
Eco mode (radio) on
Radio connection from 08:00
Radio connec-
tion until
12:00
Settings
15.03.2020 08:00
79 Free meas. heads
The settings available here
depend on the specified Type
of measurement (Standard,
BOD special, Pressure p).