Operation Multi 3630/3620 IDS
16 ba77045e01 09/2019
4.3 Settings
4.3.1 Measurement settings of the set
The measurement settings for a set are defined while the set is being estab-
lished.
Default settings are printed in bold.
Menu item Possible setting Explanation
Measuring mode BOD standard
BOD special
Pressure p
Depending on the selected measuring
mode, some of the following settings are
enabled or grayed out
Set name
max. 8 characters The 8 characters can be freely selected
Preselected is a clear combination of
numerals (date).
An index is also assigned autimatically.
Duration
30 min,1 h ...1 d... 5 d...
180d
Duration of the measurement
Settings General settings for the set
AutoTemp on
off
Function to control the temperature
adjustment. Measurement is started
automatically (at least 1 hour, at most 3
hours after the OxiTop
®
-IDS measuring
head is started).
Pre-tempering the samples to exactly
20°C is not required.
The samples can be inserted with a tem-
perature in the range of 15°C ... 21°C.
The AutoTemp function takes over the
temperature control.
Warning difference
pressure
(Only for Measuring
mode: Pressure p)
50 ...150 ... 500 hPa Difference pressure (negative or positive
pressure difference compared to the
start pressure). If it is exceeded, the
meter marks the sample with an excla-
mation mark (!).
The warning when the specified Warning
difference pressure is exceeded is trig-
gered when:
– the measurement data have been
transmitted to the meter and
– the measured value that was last
stored by the OxiTop
®
-IDS measur-
ing head exceeds the difference pres-
sure