Multi 3630/3620 IDS Operation with OxiTop
®
-IDS measuring heads
ba77045e01 09/2019 21
5.1.2 The course of the measurement
Preparation before
starting the
measurement
1 Carry out the preparatory activities
(see section 5.1.1 P
REPARING THE MEASUREMENT):
Keep ready the instruments and accessories required
Determining the sample volume
Preparing the samples
2 Switch on the meter with <On/Off>.
3 If necessary, use <M_> to switch to operation with the OxiTop
®
-IDS
measuring heads.
The OxiTop set management (overview of the sets started) is dis-
played.
During the initial commissioning the list OxiTop set management is
empty.
The display shows No data available!.
4 In the OxiTop set management, use <F1>/[New] to create a new set of
samples.
The Set menu opens.
5 Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without a temperature sensor.
190216-1 5/5 BOD5 21.05.2019
190216-2 360/360 BOD5 21.05.2019
190216-3 6/7 BOD7 23.05.2019
190217-1 4/4 BOD4h 17.05.2019
190220-1 4/20 BOD20 12.06.2019
190220-2 360/360 BOD5 21.05.2019
RED_F113-1 1/5 p5 25.05.2019
190223-1 137/360 BOD30m 23.05.2019
79 Free measuring heads
OxiTop set management
28.05.2019 08:00
DeleteNew