Firmware update Multi 3630/3620 IDS
36 ba77045d02 03/2020
8 Firmware update
8.1 Firmware update for the meter Multi 3630/3620 IDS
You can find available firmware update files for your meter on the Internet.
With the "Firmware Update " program and a PC you can update the firmware
of the Multi 3630/3620 IDS to the newest version.
For the update you have to connect the meter to a PC via the USB interface.
The update requires:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the Multi 3630/
3620 IDS).
After switching the meter off and on you can check whether the meter has taken
over the new software version
Operation with OxiTop
®
-IDS measuring heads:
Menu <ENTER_> / Service information.
When operating with IDS sensors:
Menu <ENTER_> / System / Service information.
1.
Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type), the new
data are displayed there.
2.
In the windows start menu, open the update folder and start the firm-
ware update program for the meter.
3.
Using the USB interface cable, connect the Multi 3630/3620 IDS to a
USB interface (virtual COM port) of the PC.
4.
Switch on the Multi 3630/3620 IDS.
5.
In the firmware update program, start the update process with OK.
6.
Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes up to 15 minutes. A terminatory mes-
sage is displayed after a successful programming process. The firm-
ware update is completed.
7.
Disconnect the Multi 3630/3620 IDS from the PC.
The Multi 3630/3620 IDS is ready for operation again.